Chapter 6—Alarms and Troubleshooting
Message
Purpose of Message
Temp Sensors
When the temperature trim function
unmatched
needs to change DAC by > 1 C, the
pre and post sensors are verified
against one another. This message
occurs if the two temperature sensors
are more than 0.5 C different.
Temp Control not
This message is displayed on the sign
calibrated
on screen if the temperature sensors
were not matched when they were
verified against one another.
Temperature greater than
Entered Temperature value is higher
max. value
than allowed.
Temperature has been
The operator has attempted to set a
set to min.
Temperature lower than allowed.
Test Complete
All selected self-tests passed.
Test Failed
The Alarm and/or PHT Sections of the
automated Test Sequence have failed
Testing temp sensor
In rare cases, the machine may be
put into bypass to verify the
temperature sensor. The OLC test
cannot be run at this time
TMP is High (toward 500)
The TMP has exceeded the TMP high
alarm limit value.
181
Warning! A rising TMP
may indicate a leak in the
balancing system and
should be investigated.
Type
Action Required
Warning
Press RESET key to reset alarm
The temperature trim function will be disabled until the machine is
turned off and back on.
Advisory
Calibrate the temperature control.
Dialog
The temperature will be set to the highest allowed level. Press
Message
CONFIRM to clear message and accept the maximum allowed value.
Verify that the value is acceptable.
Dialog
The temperature will be set to the lowest allowed level.
Message
Advisory
Advisory only. No action required.
Test
Reset the alarm. Check the setup to see if the alarm can be
Alarm
corrected and then retest. If the machine fails, turn machine power
Off and back On. If alarm is still not cleared, take the machine out of
service and alert a qualified service technician.
Advisory
Wait 10 minutes and start the OLC test again.
Blood
1) Check the dialyzer supply and return lines for kinks and that the
Alarm
connectors are properly connected to the dialyzer or the shunt.
2) Clean the dialysate line filter screen.
3) Press RESET key to reset alarm. Press the Override key and
hold for one second to select new alarm limits or for adjusting the
TMP. If unable to reset the alarm, call your local qualified service
technician.
High UF Goal and low dialyzer KUF coefficient can exceed the
maximum TMP of 520 mmHg. The UF Goal may need to be lowered.
This in turn will lower the UF rate and the TMP. Notify a physician if
the UF Goal has changed.
2008K Hemodialysis Machine Operator's Manual Rev P