Monitoring the Single Needle Dialysis Treatment
Follow the instructions in the "Monitoring the Treatment" section on page 95.
Warning! Administration of intravenous solution during single needle operation requires
the use of a sterile one-way valve between the administration set and the infusion site to
prevent back up of solution.
Single Needle Alarms and Troubleshooting
Follow the instructions in the "Alarms and Troubleshooting" section on page 137.
Blood Recirculation Procedure during Single Needle Dialysis
Follow the instructions in the "Blood Recirculation Procedure" section on page 117.
Power Failure during Single Needle Dialysis
Follow the instructions in the "Power Failure during Dialysis" section on page 118.
Completion of the Single Needle Dialysis Treatment
At the end of treatment, when the RTD timer has counted down to 0:00, an alarm sounds and
the message, RTD = ZERO , appears in the Status Box. An alarm also sounds when the set
amount of ultrafiltrate has been removed. When that happens, the Status Box displays the
message, UF GOAL REACHED. To reset either alarm, press the RESET key. If the UF
GOAL REACHED and RTD = ZERO alarms occur simultaneously, pressing the RESET key
will reset both alarms.
Returning Blood to the Patient
1. Select the Tx Clock button and then press the CONFIRM key to stop the treatment
2. Press the Start/Stop key on the Arterial Blood Pump module to stop the pump
3. On the "Test & Options" screen, set the Single Needle toggle-button to 'Off' (by
touching the button and pressing the CONFIRM key)
4. Remove the single needle blood pump segment from the single needle blood pump:
a. Open the door and align the rotor by pressing and holding the Start/Stop key until
the pump stops.
b. Press the clamp-panel below the rotor to release the left (incoming) side of the pump
segment. Pull the first couple of inches of the pump segment out of the pump.
2008K Hemodialysis Machine Operator's Manual Rev P
Appendix A—Single Needle Dialysis (Optional)
199