To set the 2008K hemodialysis machine for sequential
dialysis:
Figure 37 – Setting Sequential Dialysis
After 60 minutes of pure ultrafiltration, an alarm sounds and the warning message, " 60
MINUTES FLOW OFF ," appears in the Status Box. The operator has the option of
continuing pure ultrafiltration or starting dialysis. This alarm occurs only once.
To continue pure ultrafiltration, press the RESET key on the left side of the control panel.
This will silence the alarm and terminate the warning message.
To start the treatment, touch the Dialysate Flow button in the "Home" screen, set it to the
prescribed rate using the data entry keypad, press CONFIRM, and start the dialysate flow by
pressing the Dialysate Flow on/off key in the upper, right corner of the control panel. When
dialysate flow is on, the yellow light is not illuminated.
2008K Hemodialysis Machine Operator's Manual Rev P
Chapter 3—Setting Treatment Parameters
1) Set UF treatment
parameters. Set the UF
Time to include the
combined treatment and
pure UF times.
2) Turn off Dialysate
Flow. Yellow light
will flash when flow
is off.
3) Touch Dialysate
Flow to highlight it.
4) Use the
number or arrow
keys to enter 0.
5) Press CONFIRM This
will deactivate the "Flow
Off" warning for five
minutes. The Dialysate
Flow button will display
"SEQ".