162
Fail * Actuator
Temperature test has failed.
Temperature High
Fail * Actuator
Temperature Low
Fail * Actuator TMP High
Transmembrane Pressure (TMP) test
has failed.
Fail * Actuator TMP Low
Fail * Actuator Venous
Venous Pressure test has failed.
High
Fail * Actuator Venous
Low
Fail * Air Detector
Air detector test has failed.
Fail * Arterial High Soft
Arterial Pressure test has failed.
Fail * Arterial Low Soft
Fail * Arterial High Hard
Fail * Arterial Low Hard
Fail * Blood Leak 1
Blood Leak test has failed.
Fail * Blood Leak 2
Fail * Cond High Soft
Conductivity test has failed.
Fail * Cond Low Soft
Fail * (Get Neg TMP)
Get Neg TMP test has failed.
Fail * (Get Pos TMP)
Get Pos TMP test has failed.
Chapter 6—Alarms and Troubleshooting
2008K Hemodialysis Machine Operator's Manual Rev P
Test
Verify stable temp. Rerun test. If failure repeats, remove machine
Message
from service and alert a qualified service technician to recalibrate if
failure repeats
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Rerun test, if failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Reposition venous drip chamber. Rerun Test. If failure repeats,
Message
remove machine from service and alert a qualified service technician.
Test
Rerun test, if failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Verify absence of air bubbles in flow indicator. Rerun test. If failure
Message
repeats, remove machine from service and alert a qualified service
technician.
Test
Rerun test, if failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Check UF pump. Rerun test. If failure repeats, remove machine
Message
from service and alert a qualified service technician.
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.

Hide quick links: