Chapter 3—Setting Treatment Parameters
Button
90
Function
The user can choose to do the Alarm Test with this button.
If the machine is set up with an automatic test valve, the user can choose
to do the Diasafe Test with this button.
Touching the Patient ID button opens an on-screen keyboard that is used
to enter a patient's ID in the text box located to the right of the button.
The 2008K hemodialysis machine can upload treatment information to a
network database for review by clinical staff using a personal computer.
The Pediatric button activates treatment options specific to pediatric
patients. The selection is indicated by an X in the On or Off box.
The High Flux button selects parameters for the use of a high flux
dialyzer for treatment. The selection is indicated by an X in the On or Off
box.
Warning! It is essential that the 2008K hemodialysis machine
balancing system is operating properly when using high-flux
dialyzers. The machine must successfully complete a
Pressure test before treatment commences. For more
information, See "Testing the 2008K Hemodialysis Machine"
on page 52.
The Arterial Width button allows the selection of three different ranges
for the arterial pressure alarm (120, 160, and 200 mm Hg).
Note: These options will only be available if set to "User Selectable" in
the Service Mode "Options" screen.
The Venous Width button allows the selection of four different ranges for
the venous pressure alarm (100 asymmetric limits, 120, 160, and 200 mm
Hg). The asymmetric limit will close the lower venous limit after a time
delay for stabilization.
Note: These options will only be available if set to "User Selectable" in the
Service Mode "Options" screen.
Runs the UF pump at preselected UF goal and time while recirculating.
UF goal and Time are entered in the Service Mode.
The Single Needle button prepares the machine for single-needle
dialysis treatment. For more information on single-needle dialysis
treatment, see Appendix A.
2008K Hemodialysis Machine Operator's Manual Rev P