Chapter 6—Alarms and Troubleshooting
Message
Purpose of Message
Dialysate flow is off
Dialysate flow is necessary to run an
OLC test.
Dialysate flow unstable
When attempting to start an OLC test,
certain conditions are necessary,
including stable dialysate flow rate.
Dialysis
Machine is currently in dialysis mode.
Dialysis Paused
In dialysis mode, Tx clock is paused.
Dialyzer Connected?
Indicates that one of the following
conditions exist:
Diasafe Test Failed
This message advises the operator of
the status of the Diasafe self test
Diasafe Test Passed
This message advises the operator of
the status of the Diasafe self test
Diasafe Test Recovery
This message advises the operator of
the status of the Diasafe self test
E.01 (Arterial or SN
EPROM CRC error
Blood Pump Message)
E.02 (Arterial or SN
Flash ROM CRC error
Blood Pump Message)
E.03 (Arterial or SN
RAM check error
Blood Pump Message)
159
Test button selected but the
dialyzer supply and return
lines are not in the shunt.
Dialyzer supply and return
lines are on the shunt but
blood is sensed and the blood
flow is on.
Type
Action Required
Advisory
Do not attempt to run an OLC test unless the dialysate flow is set
between 300 – 800 ml/min
Advisory
Wait a minute or so and start the OLC test again.
Advisory
Advisory message only. No action is required.
Advisory
Status line advisory message only. No action is required.
Advisory
To proceed, either:
Connect the dialyzer supply and return lines to the shunt if the
procedure requires them to be connected at this time
Or,
Connect the dialyzer supply and return lines to the dialyzer if the
procedure requires them to be connected at this time.
Note: This message may also briefly appear if the blood
pump rate is set too low during setup. Raise the rate to at
least 100 ml/min when the blood pump is running.
Warning
Press the RESET key to clear the message. Rerun the test. If test
fails again return the blood to the patient if alarm occurs during
treatment. Take the machine out of service and alert a qualified
service technician to replace the Diasafe filter if necessary
Advisory
Press the RESET key to clear the message
Advisory
Advisory message only. No action is required.
Alarm
If this message occurs, in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
2008K Hemodialysis Machine Operator's Manual Rev P