Fresenius Medical Care 2008K Operator's Manual page 152

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

152
Blood Pump Stopped
The blood pump is on and the speed
is set, but the blood pump has
stopped for a period exceeding its set
time limit of either 15 or 30 seconds
(time limit is set with dip switch #4 on
the blood pump module PCB).
Blood Pump Tach Alarm
Optical tachometer not in range
Blood Pump Task Error
Software task was not completed
correctly
Blood Pump Timer Error
50 ms second time period exceeded
Blood Pump Update Error
Transmit error during Flash update
while in Service Mode
Blood Pump Volt Error
Reference Voltage error
Blood Pump WD Error
Watchdog timeout
Chapter 6—Alarms and Troubleshooting
2008K Hemodialysis Machine Operator's Manual Rev P
Blood
1) Correct other blood alarms that could have triggered the stopped
Alarm
pump message.
2) Inspect the blood pump module to see if:
The blood pump door is closed.
The pump tube segment is properly positioned. Correct if
necessary.
3) Press the RESET key to reset the alarm.
4) If running double-needle dialysis with the single needle pump in
the machine, the Single Needle option in the "Tests & Options"
screen must be off.
5) If running single-needle dialysis with the single needle pump in
the machine, the Single Needle option in the "Tests & Options"
screen must be on. Next,
Set blood flow rate to zero
Increase the blood pump rate to 100 ml/min
Check the pillow on the arterial bloodline below the arterial
blood pump for poor blood flow.
6) Slowly increase flow to the prescribed rate. If unable to resume
blood flow rate, manually return the blood to the patient if alarm
occurs during treatment. Take the machine out of service and
alert a qualified service technician to replace the blood pump
module.
Alarm
See message A.24
Alarm
See message E.15
Alarm
See message E.14
Alarm
See message E.99
Alarm
See message E.04
Alarm
See message E.06

Hide quick links: