Fresenius Medical Care 2008K Operator's Manual page 34

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Chapter 1—Overview
The Heparin Pump Module
The heparin pump provides a means of injecting heparin into the blood circuit gradually over
the course of the treatment and/or as a bolus. The pump can accommodate a variety of
syringes that are commercially available. The pump works in conjunction with the "Heparin"
screen where such parameters as the size and type of the syringe, infusion rate, infusion time,
and bolus amount of heparin to be infused are selected.
If heparin is infused manually (by pushing in the carriage lock button while pushing on the
slide carriage), the volume will not be added to the displayed amount, and must be added to
the total heparin amount. Manually moving the carriage to infuse heparin is not
recommended.
Barrel
Lock
Tabs
Wings
Slot
Slide
Carriage
Carriage
Latch
Figure 11 – The Heparin Pump Module (shown with syringe)
34
2008K Hemodialysis Machine Operator's Manual Rev P
Syringe
Syringe
Plunger

Hide quick links: