Available Software & Hardware Treatment Options and
Default Settings
The following options are available on the 2008K hemodialysis machine and selected from
the Service Mode in either the "Treatment Options" screen or "Hardware Options" screen.
Forced Test
Upon power up and entering dialysis (except after a power failure or short power down), the
machine will be in Standby. In Standby, the blood pump doesn't run (except for Prime and
Level Adjust). The ultrafiltration and Sodium Variation System cannot be turned on.
"STANDBY FOR TEST" is displayed as a low priority message.
The test sequence will automatically start 30 seconds after the test criteria have been met.
Standby ends with the initiation of the test (whether by the operator or automatically).
Spread Limits
When activated and no blood leak alarm exists, the Override key can be used to spread the
arterial and venous alarm limits by 300 mmHg for 30 seconds. The TMP alarm limits will
completely open. After 30 seconds the limits will reset around the current pressure readings.
Auto Blood Pressure Reading
This option can be used to select the method for determining when to take a blood pressure
reading. With "Interval" selected, the readings occur at the time interval selected. With
"Clock Time" selected, the blood pressure is taken at specified times (e.g. every half hour on
the hour and half hour).
Auto Heat Disinfect
In the "Options: Default Settings" screen the user may select a start time (Start Prg) and days
of the week toggle-buttons, and a pre-rinse time to automatically run a Heat Disinfect
program. This program affects only the start time of the Heat Disinfect, all other settings
function as usual. With the machine on and set up for rinse, the Heat Disinfect program will
run at the selected time(s).
Note: Requires special activation for this feature, contact Fresenius Medical Care Spare Parts
department or your sales representative for more information and pricing.
Off After Heat Disin
Set in the "Options: Default Settings" screen, the user may choose to automatically power
down the machine after an Auto Heat Disinfect program is run.
2008K Hemodialysis Machine Operator's Manual Rev P
Appendix B
211