Chapter 4—Monitoring the Treatment
The blood pressure readings are displayed both in table and graph form (The graph can also
be viewed in the "Trends" screen). The table lists the time the blood pressure reading was
taken, the systolic and diastolic pressures, the Mean Arterial Pressure (MAP), and the pulse
rate of the patient during the test. The MAP is measured by the blood pressure module and
thus may differ from MAP calculated from systolic and diastolic pressure.
The pressure readings on the graph are represented by vertical lines with ticks at the top and
bottom signifying the systolic and diastolic pressures respectively. The first pressure reading
is displayed on the left side of the graph with subsequent readings appearing to the right. The
table on screen displays a maximum of 10 pressure readings at a time.
first one. The results of both automatic and manual tests are displayed in the table. Results
appear in the graph only after the Tx Clock is started.
Note: For accurate blood pressure readings, the cuff must be the proper size and positioned
at heart level. Each centimeter above or below the heart that the cuff is positioned, will
result in a reading error of 0.8 mm Hg. Tests taken with the Tx Clock paused do not show
up on graph.

During Treatment

Online Pressure Holding Test
The online Pressure Holding Test (PHT) automatically checks the integrity of the dialysate
balancing system during dialysis when the dialyzer is connected. The online PHT detects
most leaks in the hydraulics that would affect the precise volumetric control of fluid in the
dialysate system.
The online PHT complements the self-test sequence; it is not a substitute. It is still necessary
to perform the initial Automatic Test Sequence before each high flux treatment.
The online PHT runs every 12 minutes regardless of the other alarm conditions. Dialysate
flow must be on and the machine cannot be executing a filling program or OLC test. The test
runs for two balancing-chamber cycles (about seven seconds). The message " RUNNING
ONLINE PHT " displays during the test. Before the test, the UF pump stops in the middle of
a cycle and remains off during the two balancing chamber cycles of the online PHT. The UF
green light will flash during this time. The machine is in bypass mode during the test period.
The displayed TMP during this time represents pressure within the hydraulics, therefore, the
TMP reading may change slightly. The TMP alarm limits are spread during the test.
116
The Stat/Deflate key, located on the right side of the control panel, can be
used to quickly relieve the pressure from an inflated blood pressure cuff. It
will also start an unscheduled blood pressure reading if the cuff is deflated.
Unscheduled tests do not have any effect on the scheduled tests. For example,
if the tests were scheduled at 15-minute intervals, and a manual test was taken
five minutes after the first test, the next test will still occur 15 minutes after the
2008K Hemodialysis Machine Operator's Manual Rev P