Fresenius Medical Care 2008K Operator's Manual page 144

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

144
60 Minutes Flow Off
Dialysate flow has been off for 60
minutes in the sequential option.
A.11 (Arterial or SN
Pump is not reaching speed at
Blood Pump Message)
maximum voltage
A.13 (Arterial or SN
Pump is turning in the wrong direction
Blood Pump Message)
A.16 (Arterial or SN
Key stuck or held in too long
Blood Pump Message)
A.20 (Arterial or SN
Set speed-read back analog voltage
Blood Pump Message)
at X348/14 is out of limits
A.21 (Arterial or SN
Actual speed-read back analog
Blood Pump Message)
voltage at X348/10 is out of limits
A.22 (Arterial or SN
Arterial pressure-read back analog
Blood Pump Message)
voltage at X348/7 is out of limits
SN pressure-read back analog
voltage is out of limits
A.24 (Arterial or SN
Optical tachometer not in range
Blood Pump Message)
Chapter 6—Alarms and Troubleshooting
2008K Hemodialysis Machine Operator's Manual Rev P
Warning
1) Press RESET to silence the alarm.
2) Re-establish dialysate flow to comply with the prescribed
treatment. The machine will go into bypass mode until dialysate
temperature and conductivity settings are attained (about two
minutes). Dialysate flow must be re-established for a minimum of
five minutes before resuming pure ultrafiltration (UF) or warning
will reoccur.
Alarm
Press the RESET key to clear. If problem persists, return the blood
to the patient if alarm occurs during treatment. Take blood pump
module out of service and alert a qualified service technician.
Alarm
Press the RESET key to clear. Verify pump rotor is turning in a
clockwise direction. If not, manually return the blood to the patient if
alarm occurs during treatment (see page 118 for instructions). Take
blood pump out of service and alert a qualified service technician.
Alarm
Press the RESET key to clear. Verify when adjusting settings, the
operator does not hold the key too long. If problem persists, return
the blood to the patient if alarm occurs during treatment. Take blood
pump module out of service and alert a qualified service technician.
Alarm
Press the RESET key to clear. If problem persists, return the blood
to the patient if alarm occurs during treatment. Take blood pump
module out of service and alert a qualified service technician.
Alarm
Press the RESET key to clear. If problem persists, return the blood
to the patient if alarm occurs during treatment. Take blood pump
module out of service and alert a qualified service technician.
Alarm
Press the RESET key to clear. If problem persists, return the blood
to the patient if alarm occurs during treatment. Take blood pump
module out of service and alert a qualified service technician.
Alarm
Press the RESET key to clear. If problem persists, return the blood
to the patient if alarm occurs during treatment. Take blood pump
module out of service and alert a qualified service technician.