Fresenius Medical Care 2008K Operator's Manual page 157

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Chapter 6—Alarms and Troubleshooting
Message
Purpose of Message
Conductivity Limits set to
Conductivity limits were found to be
default
outside of the allowed range at power
up.
Conductivity Low
The actual or measured conductivity
has exceeded the low conductivity
alarm limit. The machine is in bypass
mode.
157
Type
Action Required
Opening
After entering Dialysis, verify that the conductivity limits are as
Screen
desired.
Dialysate
1) Check to see if :
Alarm
Dialysate flow is on.
The correct concentrate is selected in the "Dialysate" screen
and the concentrate supply lines are connected to
appropriate concentrate sources.
The prescribed concentrate and the correct baseline Na+ and
Bicarbonate values are displayed in the "Dialysate" screen.
The supply of concentrate is adequate.
The concentrate has been mixed properly, (i.e. bicarbonate
mixed well with RO water).
2) Verify that the concentrate connectors are sucking concentrate. If
not:
Turn off dialysate flow and disconnect the concentrate
suction connectors from their wands.
Check for clogged filter screens in the connector handles,
especially the bicarbonate connector. Clean if necessary.
Re-assemble the concentrate connectors. Verify that the
connectors and filter assemblies are tightly screwed together
with no air leak.
Check that the O rings on the tips of the concentrate
connectors are not damaged or missing.
Reconnect the connector to the concentrate source. Turn on
dialysate flow and recheck the connectors for suction. If
suction is present, allow 5 minutes for conductivity to reach
the prescribed level.
If suction is not present in both connectors, discontinue treatment and
remove patient from the machine. Perform an Acid Clean program
followed by a complete rinse cycle. Test machine operation. If
conductivity alarm persists, take the machine out of service and alert
a qualified service technician.
2008K Hemodialysis Machine Operator's Manual Rev P

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents