5.4
Access
Description
LS10/LS15, Measurement Guidelines
Manual Input Screen for Optical Height Reading
The Manual Input screen can only be accessed from within a levelling applica-
tion.
1.
Press the Favourites key.
2.
Change to the Apps tab.
3.
Select
Man.Input.
If electronic height readings are not possible, you can perform an optical height
reading and use the Manual Input screen to enter the measured data manually.
With the left and right navigation keys select one of the desired distance methods:
Distance Method Optical
Distance Method Tape
Perform an optical height reading and enter
the values for the staff height and for the
heights of the distance line above and
below. The instrument automatically calcu-
lates the distance based on the entered
staff section and the multiplication constant
100.
Press Cont to add the entered data to the
current levelling line.
Perform an optical height reading and
measure the distance with a tape.
Enter the values for the staff height and for
the distance.
Press Cont to add the entered data to the
current levelling line.
33