Adjustment Method
"A x Bx"
LS10/LS15, Check & Adjust
Step
Description
An information page displays the correct setup of the station according to the
6.
selected adjustment method.
A x x B method
Set up the instrument at the second position (station 2) and level the instru-
ment with the electronic level.
7.
Press F4 or Cont.
8.
• Aim at staff B and carry out a measurement (B2).
• Aim at staff A and carry out a measurement (A2).
• Store the measurements for station 2.
On the page
Result
9.
For some markets, it may be necessary to add the determined user
collimation error to the factory collimation error. By adding the user
collimation error, the correct collimation error is applied to all meas-
urements and the reported user collimation error is within the market-
specific tolerance.
To add the determined user collimation error to the factory collimation
error, press Mve Er. You need to confirm this action by pressing Cont.
After confirming, the new user collimation error is reported as 0.0.
"From the Centre" Procedure
A
1/2
1/2
1
a
009364_001
First position (station 1):
• Position the instrument centrally between staff A and B.
• The centre has to be within an accuracy of ±1 m.
• The distance a between the staffs should be approximately 30 m.
Second position (station 2):
• Position the instrument close to staff B (inside or outside).
• The distance b has to be at least 2.5 m.
A x Bx method
, the results of the adjustment procedure are displayed.
To set the determined collimation error and
apply it to all subsequent measurements,
press Set.
B
b
2
b
88