Leica LS10 User Manual page 112

Digital level
Hide thumbs Also See for LS10:
Table of Contents

Advertisement

CONF Commands
LS10/LS15, GSI Online Commands
Function
Protocol
Data Recording
Delay (between
2 strings sent)
AutoOff
Earth Curvature
Correction
Staff Mode
Output GSI Format
Length
Code Recording
Syntax
CONF/<conf spec>
<CR/LF>
Function
Spec. Command
Beep
30
Display Illumina-
32
tion
Unit (Length)
41
Unit (Tempera-
42
ture)
Spec.
Setting
75
0 = OFF
1 = ON
76
0 = Internal Memory
1 = RS232
78
Range from 0 to 50:
0 = No delay
25 = 25 ms delay
50 = 50 ms delay
95
0 = Disable
1 = Enable
2 = Standby
125
0 = Off
1 = On
127
0 = Upright
1 = Inverted
137
0 = GSI-8
1 = GSI-16
138
0 = Before measure-
ment
1 = After measure-
ment
Response
00<conf spec>/00 <param-
eter>
Response
CONF/30
0030/0000
0030/0001
0030/0002
CONF/32
0032/0nnn
CONF/41
0041/0000
0041/0001
0041/0002
CONF/42
0042/0000
0042/0001
Example
SET/75/1
Setting Acknowledgement
protocol to ON
SET/76/1
Setting data recording to
external via RS232
SET/78/25
Setting delay to 25ms
between subsequent strings
SET/95/1
Setting AutoOFF to ON
SET/125/1
Setting Earth Curvature
Correction to ON
SET/127/1
Setting Staff to inverted
mode
SET/137/1
Setting format output to
GSI-16
SET/138/1
Setting Code recording after
the measurement
Example
READING BEEP SETTING ON
LS10/15:
Command: CONF/30
Response: 0030/0002
Parameters
0 = OFF
1 = Medium
2 = Loud
Range from 0%-100%
0 = 0% contrast
50 = 50% contrast
100= 100% contrast
0 = Meter
1 = US ft, decimal
2 = International ft,
decimal
0 = °C
1 = °F
112

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ls15

Table of Contents