5.10 Performing inhalation
Caution!
The device must not be operated in combination with a nebulizer during oxygen in-
halation via an inhalation hose system. Otherwise the device does not generate suf-
ficient pressure for the ventilation and thus inhibits the therapy.
If you have a unit equipped with CO
can vary the display as follows:
– Pressure, flow
– Pressure, CO
– Pressure, flow, CO
– Pressure, flow, measurements
– Pressure, CO
– Gauge (only in volume-controlled modes)
1. Connect the inhalation adapter to the ventilation hose
port on the unit.
or
Connect the inhalation adapter to the patient-side port
of the patient hose system.
2. To block the measuring ports on the device during the
inhalation, place the cover on the inhalation adapter on
the upper two measurement ports on the device.
Notice
The cover is not required when connecting the in-
halation adapter to the patient-side port of the pa-
tient hose system. In this case, the connection plug
of the measuring hose system blocks the measur-
ing ports instead.
3. Switch on the device.
4. Select "New patient".
measurement, you
2
2
2
, measurements
2
Operation
EN
53