Weinmann MEDUMAT Transport Instructions For Use Manual page 42

For devices from software version 6.7. ventilator
Hide thumbs Also See for MEDUMAT Transport:
Table of Contents

Advertisement

EN
Operation
42
The self-test is successful when all the steps have been
completed. Check that all the steps are successfully
completed. Do not operate the unit if:
• one of the first three steps has not been
successfully completed
• the last part has not been successfully completed
("Fault" message appears in the display)
Caution!
The automatic self-test is not a substitute for a function
check. Before using the unit, always carry out a
function check as described in section"10. Function
check"on page 115. This is the only way to ensure that
the unit is fully functional.
4. The "Start menu" appears in the display. You now have
the following options:
– Press one of the emergency buttons (Infant, Child,
Adult): The unit immediately begins ventilation
with the preset parameters.
– Press the "Previous patient" button: The
ventilation settings from the previous patient
appear. Select the appropriate parameters for the
ventilation of the current patient.
Notice:
– If the switch-off time of the device is less than
30 seconds, the device starts ventilation
automatically after a 20-second countdown.
– Press the "New patient" button: Select the
"Adult", "Infant" or "Child" setting. The "Mode"
menu appears. Select the appropriate ventilation
mode and confirm your selection. Use the control
knobs to set the parameters for ventilating the
patient.