Weinmann MEDUMAT Transport Instructions For Use Manual page 71

For devices from software version 6.7. ventilator
Hide thumbs Also See for MEDUMAT Transport:
Table of Contents

Advertisement

You can find more setting options under the menu item "Advanced ventilation parame-
ters" in the "Main menu" (see "7.4 Advanced ventilation parameters" on page 87).
Pressure
pMax
Automated
ventilation
pInsp
Pressure
dT
PEEP
The aPVC mode is used for pressure-controlled, assisted ventilation at a fixed mandatory
ventilation rate.
In case of spontaneous respiration, the patient has the possibility of increasing the rate and
consequently the minute volume MV. If the patient displays a spontaneous respiratory ef-
fort within a specified time slot of the expiration, the mandatory mechanical breath is syn-
chronized with the patient's respiration. The time slot or trigger time slot can be set in %
of Te before the next expected mandatory mechanical breath. If the patient displays a
spontaneous respiratory effort outside of the set trigger time slot, no mandatory mechan-
ical breath is triggered.
Synchronized
automated
ventilation
ramp
dP
1/Freq. (current)
1/Freq. (set)
Synchronized time slot
(adjustable)
ΔT
1/Freq. (set)
Ventilation modes
Automated
ventilation
Time
EN
71