Weinmann MEDUMAT Transport Instructions For Use Manual page 80

For devices from software version 6.7. ventilator
Hide thumbs Also See for MEDUMAT Transport:
Table of Contents

Advertisement

S-IPPV
S-IPPV: Synchronized Intermittent Positive Pressure Ventilation
Warning!
• Risk of hyperventilation! Continuously monitor the patient's measured
respiratory rate and measured minute volume in order to prevent
hyperventilation. To this end, set a narrow alarm limit for the
to recognize the risk of hyperventilation in good time.
• Risk of air trapping! Continuously monitor the airway pressure in order to
prevent air trapping.
• Risk of intrinsic PEEP! An expiration that is too short can cause the pressure to
increase slowly at the end of the expiration. Monitor the sensitivity of the
inspiratory trigger. Once the set PEEP is exceeded, a high-priority alarm (
guarantees the safety of the patient.
You can set the following ventilation values using the control knobs:
Ventilation
Control
mode
knob 1
PEEP
S-IPPV
EN
Ventilation modes
80
1
2
3
Control
Control
knob 2
knob 3
Vt
pMax
4
5
Control
Navigation
knob 4
knob 5
I:E and
Freq.
Select/
Confirm
alarm, in order
f
Function
Function
button 6
button 7
-
Trigger
)
PEEP
8
7
6
Function
button 8
Mode

Advertisement

Table of Contents
loading

Table of Contents