You can find more setting options under the menu item "Advanced ventilation
parameters" in the main menu (see "7.4 Advanced ventilation parameters" on page 87).
Pressure
synchronized automated
automated ventilation
ventilation
pMax
pPlat
PEEP
Time
ΔT
1/Freq. (current)
1/Freq. (set)
Synchronized time slot
The S-IPPV mode is used for volume-controlled ventilation with a variable mandatory
minute volume MV. Throughout the entire expiratory phase, a trigger is active which
enables the patient to trigger a new breath. This means the patient has the option of
increasing the respiratory rate and therefore the minute volume MV, and adapting these
to his/her needs. As a rule this mode is used on patients who have inadequate spontaneous
respiration.
Ventilation in the S-IPPV mode corresponds to ventilation in the IPPV mode with the
difference that it is possible to synchronize ventilation with the patient's efforts to inhale.
Since the setting for the respiratory rate is lower, the patient can trigger mandatory
mechanical breaths spontaneously. A trigger time slot extending throughout the expiration
time is available for this synchronization.
Ventilation modes
EN
81