Display of alarms
Info field
Muting the alarm
MEDUMAT Transport displays alarms as follows:
• High priority
– LED flashes red
– "High priority" alarm sounds every 8 seconds
– Alarm text appears in info field; info field flashes
red
– Corresponding alarm limit in the measurement
field flashes red
• Medium priority
– LED off
– "Medium priority" alarm sounds every 15 seconds
– Alarm text appears in info field; info field flashes
yellow
– Corresponding alarm limit in the measurement
field flashes yellow
• Low priority
– LED off
– "Low priority" alarm sounds every 30 seconds
– Alarm text appears in info field; info field has
turquoise background
Set alarms are displayed in the "Alarm Limits" menu in
their priority color (see "7.2 Alarm Limits" on page 85).
You can use the alarm mute button at any time to mute
the alarm signal for 120 s. During this time, the yellow LED
of the alarm mute button flashes. Pressing the button
again reactivates the acoustic alarm. The visual alarm
continues to remain active.
Visual and acoustic alarms are reset automatically as soon
as the cause of the alarm has been removed.