Dust load on filter
Suction filter
Allow 0% audio volume
Allow alarm LED off
Allow NVG
EN
Operator menu
108
This menu item enables you to set the hygiene input filter
wear according to the level of contamination in the ambi-
ent air. You can choose between the settings "normal",
"high", and "very high". Select "normal" for standard
conditions, and "high" or "very high" for very dusty or
contaminated ambient air. This menu item is only visible if
you have activated the hygiene input filter in the menu
item "Hygiene input filter installed".
This menu item enables you to indicate that you changed
the suction filter before the filter service life expired. Doing
so resets the suction filter counter. This menu item is not
available if you activate the hygiene input filter in the op-
erator menu.
This menu item allows you to set the alarm volume to
<50% in the menu item "Audio/Video" in the main menu
(see "7.6 Audio/Video" on page 90). This menu item is
only visible if you have enabled the NVG option in the op-
erator menu (see "8.5 Options" on page 105).
This menu item allows you to activate/deactivate the alarm
LED in the menu item "Audio/Video" in the main menu
(see "7.6 Audio/Video" on page 90). This menu item is
only visible if you have enabled the NVG option in the op-
erator menu (see "8.5 Options" on page 105).
This menu item allows you to activate the "NVG" menu
item in the main menu (see "7.9 NVG (Night Vision Gog-
gles)" on page 94). This menu item is only visible if you
have enabled the NVG option in the operator menu (see
"8.5 Options" on page 105).