Weinmann MEDUMAT Transport Instructions For Use Manual page 36

For devices from software version 6.7. ventilator
Hide thumbs Also See for MEDUMAT Transport:
Table of Contents

Advertisement

Assembling the nebulizer
Caution:
When assembling the nebulizer, observe the correct order of the individual
components. Always install an HME filter, a bacteria filter, or a combined HME
bacterial filter between the BiCheck flow sensor and the nebulizer. If the filter is not
installed correctly or no filter is used, the membranes in the patient valve may
become stuck together and cause the BiCheck flow sensor to deliver faulty
measurements.
1. Attach the mask/tube (with elbow if desired) to the nebulizer.
2. Connect the open end of the nebulizer to the HME filter, bacteria filter or the combined
HME bacterial filter.
3. Attach the HME filter, bacteria filter, or combined HME bacterial filter to the BiCheck
flow sensor of the patient hose system.
Supplying oxygen to external units
You can use the O2/AIR inlet/outlet to connect units, modules or inhalation devices to the
MEDUMAT Transport (quick connector to the front of the units).
When doing so, bear in mind that the outlet gas flow reduces the efficiency of the gas
supply (see "14.6 Possible O
EN
Installation
36
concentration with counterpressure" on page 149).