Weinmann MEDUMAT Transport Instructions For Use Manual page 62

For devices from software version 6.7. ventilator
Hide thumbs Also See for MEDUMAT Transport:
Table of Contents

Advertisement

Ending battery charging
MEDUMAT Transport automatically determines the optimum point at which to end
charging by measuring and evaluating the charging curve and battery temperature. As
soon as charging has ended, the charging indicator glows green continuously.
Interrupting battery charging
Battery charging is continuously monitored by MEDUMAT Transport.
Faults which cause charging to be interrupted are:
Battery temperature rises above 45°C, e.g., due to high ambient
temperature, or falls below +5°C
Charging current is too high (>3 A) e.g., due to a short-circuit
If one of these problems occurs, charging is automatically interrupted and the charging
indicator glows red.
Charging is also interrupted if there is no external power supply, e.g., because the unit,
which is mounted on a portable system, is removed from the wall mounting. The charging
indicator does not light up in this case because this is not a fault, but a normal operating
situation. As soon as the external power is restored, e.g., when the portable system is put
back into the wall mounting, charging continues immediately.
EN
Operation
62

Advertisement

Table of Contents
loading

Table of Contents