For devices from software version 6.7. ventilator
Hide thumbs Also See for MEDUMAT Transport:
Table of Contents

Advertisement

You can activate the NVG mode here. When the NVG
mode is activated, the device behaves as follows:
• Alarm LED deactivated
• Acoustic alarm output for all alarms permanently deac-
tivated
• Coloring of the display optimized for night vision devices
symbol in display and
fields
• Display brightness reduced as per preset (see "7.6 Au-
dio/Video" on page 90)
This menu item only appears if you enable the NVG option
in the operator menu (see "8.5 Options" on page 105).
This option is only permitted for use in the military sector.
A device in NVG mode does not comply with the following
standards with respect to alarm output:
• EN 60601-1-8
• EN 794-3/EN 10651-3
The operator assumes the resulting risk for operation.
symbol in measurement
Main menu
EN
95