– Press the "New patient" button: Select the
"Height" setting. The "Patient parameters" menu
appears. Select the gender. Set the correct height
with the context-dependent function keys or the
navigation knob. Confirm the setting with
"continue". Now select the appropriate ventilation
mode and confirm your selection. If necessary, use
the control knobs to change the parameters for
ventilating the patient (see "14.8 Calculation of
body weight on the basis of body height" on
page 152).
Then use the context-dependent buttons to select
"Start ventilation" if you want to start ventilation,
"Alarm Limits" if you want to determine the alarm
limits in the "Alarm Limits" menu (see "7.2 Alarm
Limits" on page 85) or "back" if you want to
change any settings.
Notice:
As soon as you select a new patient, you can use
the menu button to change over to the main
menu.
– Select "Function check" menu: The unit will begin
the automatic function check (see "10.4
Performing a function check" on page 119).
Notice:
Following the start of ventilation, all alarms are
automatically muted for 120 seconds. This is with
the exception of the technical alarms "Supply
pressure < 2.7 bar", "Battery almost empty", and
"Device malfunction", which cannot be muted.
During this time, visual alarms are still displayed.
5. When the self-test has finished and the ventilation
mode has been set, connect the patient.
6. Adjust the ventilation values if necessary during
ventilation.