Weinmann MEDUMAT Transport Instructions For Use Manual page 33

For devices from software version 6.7. ventilator
Hide thumbs Also See for MEDUMAT Transport:
Table of Contents

Advertisement

Elbow
Hose protection sleeve
Water filter for CO
Notice:
Always operate the unit with a water filter when CO
particles that have been sucked in can damage the CO
The water filter WM 97012 loses efficiency after approx. 8 hours of continuous operation,
depending on the temperature, humidity and any coarse particles such as mucus.
Change the filter after eight hours at the latest.
The filter's decreasing efficiency is indicated by the alarm message "CO
display. This message is accompanied by a medium-priority audible alarm.
Notice
You can remove the elbow to reduce the dead space or
to adapt the hose routing to suit the patient's position.
The tube protection sleeve is pulled over the ventilation
hose with connected BiCheck flow sensor. It prevents the
hose system from tangling on other items of equipment
and being damaged.
measuring hose
2
suction is activated. Otherwise
2
module.
2
occlusion" on the
2
Installation
EN
33