Pre-oxygenation
Inhalation
You can activate or deactivate apnea ventilation in the
"Apnea ventilation parameters" menu. The "Apnea
mode" menu item allows you to select which ventilation
mode the unit utilizes during apnea ventilation:
• BiLevel + ASB (pressure-controlled)
• SIMV + ASB (volume-controlled)
In this menu, you can also set the ventilation parameters for
Apnea ventilation. If you do not set any ventilation parameters,
the unit adopts the presettings for these ventilation modes
according to the patient (Infant, child, adult).
Pre-oxygenation is a function which enables the oxygen
content in a patient's lungs to be increased quickly, e.g., in
preparation for intubation.
The device administers a flow of 10-25 l/min with 100%
O
, which can be set in the menu item "Pre-oxygenation".
2
Pressure is limited to 10 mbar for safety reasons.
You can select pre-oxygenation in the "Mode" menu (see
"6. Ventilation modes" on page 63). To deactivate pre-ox-
ygenation, switch to another ventilation mode.
Inhalation is a function which makes it possible to increase
the oxygen content in a patient's lungs. The inhalation
flow can be set at a value between 0 l/min and 10 l/min.
The unit administers a flow of 100% O
in the menu item "Inhalation".
You can select inhalation in the "Mode" menu (see "6.
Ventilation modes" on page 63).
Caution!
In the case of oxygen inhalation, check that the inhalation
hose is connected and the measuring ports on the
MEDUMAT Transport are blocked (see "4.3 Connecting
the inhalation adapter" on page 34).