Operator menu
100
Notice:
If a ventilation mode is already deactivated
cannot call up its presets.
If you have selected and confirmed the "edit presets" field,
you enter the "Edit ventilation settings" submenu for the
required ventilation mode. This submenu is divided into
the patient groups Infant, Child, and Adult.
Further information on the possible settings can be found
in the "Advanced ventilation parameters" sections (see
"7.4 Advanced ventilation parameters" on page 87) and
"Technical data" (see "14. Technical Data" on page 142).
The non-selectable functions in a particular ventilation
mode are not shown.
5. To switch between the "edit ventilation settings" and
"edit alarm limits" submenus, press the function key in
the top left (see image).
6. To reset the alarm limits to the factory settings, select
the "reset alarm limits" field in the column of the re-
spective patient group and confirm your selection.
Further information on the alarm limits can be found in
Section "7.2 Alarm Limits" (see "7.2 Alarm Limits" on
page 85).
7. To reset the ventilation settings to the factory settings,
select the "reset vent. settings" field in the column of
the respective patient group and confirm your selec-
tion.