Weinmann MEDUMAT Transport Instructions For Use Manual page 122

For devices from software version 6.7. ventilator
Hide thumbs Also See for MEDUMAT Transport:
Table of Contents

Advertisement

Component
Pneumatic
Sensors
Key and rotating switch test
Suction filter
or
Hygiene input filter
If, despite all your measures to rectify the faults, a component still remains marked with a
red cross, please contact an authorized dealer or WEINMANN Emergency.
EN
Function check
122
A red cross next to "Pneumatic" and "Sensors" can indicate
a secondary fault; for this reason, correct the "BiCheck",
"Hose system" and "Suction filter" faults first and then
repeat the function check. If, after that, "Pneumatic" and
"Sensors" are still marked with a red cross, have the unit
serviced by an authorized specialist dealer or by
WEINMANN Emergency.
Have the unit serviced by an authorized specialist dealer or
by WEINMANN Emergency.
Change suction filter.
or
Change hygiene input filter
Rectification