5.6 Changing the ventilation mode

To change the currently set ventilation mode, proceed as follows.
If you change from one ventilation mode to another, the unit will respond as follows:
Ventilation parameters which are also available in the new ventilation
mode are retained unchanged.
Ventilation parameters which are not available in the new ventilation
mode are saved, but have no influence on current ventilation progress.
EN
Operation
48
1. First, use the "Mode" function button to select the
"Mode" menu.
2. Then use the navigation knob or the context-depen-
dent function buttons on the right of the display to se-
lect whether you wish to ventilate invasively or non-
invasively.
The device then proposes the possible ventilation
modes to you. The possible ventilation modes for non-
invasive ventilation are: BiLevel + ASB, aPCV, PCV and
CPAP + ASB.
3. Then select the required ventilation mode. Alternative-
ly, you can select the "Inhalation" or "Pre-oxygen-
ation" function.
4. Confirm your selection by pressing the navigation knob
or the corresponding context-dependent function but-
ton.
5. Set the ventilation parameters for the selected mode.
6. Then select "continue" to proceed to the "Advanced
ventilation parameters" submenu and confirm your se-
lection.
Continue from point 7.
or
Select "Start ventilation" and confirm your selection.
The ventilation begins in the newly selected ventilation
mode.
7. Set the advanced ventilation parameters for the select-
ed mode.
8. Select "Start ventilation" and confirm your selection.
The ventilation begins in the newly selected ventilation
mode.