10. Function check

The user must carry out a function check on the unit before each use and after each
disassembly, but at least every 6 months.
Warning!
If this function check reveals any faults or deviations from the specified values, you
must not use the MEDUMAT Transport. Have the unit repaired by WEINMANN
Emergency or an authorized dealer.
You should first try to rectify the fault with the aid of the information provided in section
"11. Troubleshooting"on page 123. If this is not possible, have the unit repaired by the
manufacturer WEINMANN Emergency or by a technician specifically authorized by the
manufacturer.
A full function check comprises:
Visual inspection for mechanical damage
Visual inspection of the display
Visual inspection of the hygiene input filter for mechanical damage and
soiling
"10.2 Checking the system for leaks"on page 117
"10.3 Checking the patient valve (reusable hose system only)"on
page 118
"10.4 Performing a function check"on page 119
We recommend that you always keep the following spare parts available:
Notice:
Carry out a visual inspection of the test lung. The test lung must not be damaged.
– Spare seals for the unit connections
– A replacement dust filter
– PEEP control diaphragm for patient valve
– Check-valve diaphragm for patient valve
– Measuring tube system, including water filter, for CO
measuring
Function check
EN
2
115