Weinmann MEDUMAT Transport Instructions For Use Manual page 144

For devices from software version 6.7. ventilator
Hide thumbs Also See for MEDUMAT Transport:
Table of Contents

Advertisement

Inspiration time:
Tidal volume
Ventilation pressure
Pressure support (Δ pASB)
Mechanical safety valve
PEEP
Trigger
Trigger sensitivity with
assisted ventilation (flow
trigger)
Trigger time slot in aPCV
ventilation mode
Flow:
- Pre-oxygenation
- Inhalation
Vt/kg body weight
Tolerances for respiratory
time ratio
Tolerances for ventilation
rate
Tolerances for tidal
volume
Tolerances for ventilation
pressure
Tolerances for flow
measurement
(monitoring)
Tolerances for volume
measurement
(monitoring)
CO
measurement
2
EN
Technical Data
144
(With pressure-controlled ventilation modes: pressure control via
With volume-controlled ventilation modes: pressure limitation to pMax
Pressure limitation to 100 mbar max.
0 mbar - 30 mbar (pressure control via proportional valve)
Flow trigger, recording of the flow value via internal flow sensor
Inspiration trigger: 1 l/min - 15 l/min
Expiration trigger: 5% - 50% of the inspiratory maximal flow
4 ml/kg body weight -10 ml/kg body weight
Side-stream method, removal rate 80 ml/min
Notice: CO
measuring is cross-sensitive to oxygen and displays a value
2
10% less than the measurement during ventilation with 100% oxygen.
MEDUMAT Transport
min. 0.2 s (200 ms)
max. 59 s
50 ml - 2000 ml
3 mbar - 60 mbar
proportional valve)
0 mbar - 30 mbar
0%-100% of Te
10 l/min-25 l/min
0 l/min-10 l/min
± 5%
-1
± 1/min
± 20 ml or ± 15%
± 3 mbar or ± 10%
± 20%
± 20%