Weinmann MEDUMAT Transport Instructions For Use Manual page 124

For devices from software version 6.7. ventilator
Hide thumbs Also See for MEDUMAT Transport:
Table of Contents

Advertisement

Fault
Alarm LED flashes,
acoustic signal sounds,
screen is dark
No acoustic alarm output
Alarm LED does not light
up
Display too dark
EN
Troubleshooting
124
Cause
Battery empty and device not
connected to the line power.
Battery defective and device not
connected to the line power.
No battery inserted and device
not connected to the line power.
Internal error
NVG mode activated
Brightness of the display set too
low
NVG mode activated
Remedy
Recharge battery or connect device to
the line power.
Use new battery or connect device to
the line power.
Insert battery in device or connect
device to the line power.
Switch the MEDUMAT Transport off
and back on again. If the error persists:
Have device repaired.
Deactivate NVG (see "NVG (Night
Vision Goggles)" on page 94).
Increase brightness of the display (see
"Audio/Video" on page 90)
Deactivate NVG (see "NVG (Night
Vision Goggles)" on page 94).