7. Main menu

7.1 Activating automatic alarm limits

General
EN
Main menu
84
In the main menu, you can optimize the unit's settings to
suit the particular operating conditions. The main menu
can be called up at any time using the "Main menu"
function button.
To navigate in the menu, you can use either the navigation
knob or the context-dependent function buttons on the
right of the display (see "5.1 Controls" on page 38).
When the "Activate automatic alarm limits" function is
activated, the unit automatically determines limits for the
alarms relating to respiratory physiology. Alternatively, you
can activate the automatic alarm limits by keeping the
alarm mute button depressed (approx. 2 seconds).
Crucial for determining the limits are the respiratory values
measured at the moment when the function is activated
(Vt
, MV
, f, if necessary etCO
e
e
In the "Alarm Limits" menu (see "7.2 Alarm Limits" on
page 85), you can set the amount of deviation (in %) from
the current respiratory value, which should trigger an
alarm.
Notice:
This overwrites the values set in "Alarm Limits" (see
"7.2 Alarm Limits" on page 85).
).