Weinmann MEDUMAT Transport Instructions For Use Manual page 112

For devices from software version 6.7. ventilator
Hide thumbs Also See for MEDUMAT Transport:
Table of Contents

Advertisement

Reusable components
Parts
MEDUMAT Transport
BiCheck flow sensor lead
BiCheck flow sensor
Reusable patient valve
Reusable ventilation
mask
Reusable ventilation hose
Reusable measuring hose
system
– PEEP control hose
– Pressure-measurement
hose
– Connector
measuring hose system
Blanking plug
Protective cap
Reusable tube protection
sleeve
EN
Hygienic preparation
112
Cleaning
Wipe down with a
moist cloth
Wipe down with a
moist cloth
In hot water with a
(5)
mild household
detergent
In hot water with a
mild household
detergent
In hot water with a
mild household
detergent
In hot water with a
mild household
detergent
In hot water with a
mild household
detergent
– Wipe down with a
moist cloth: Using
water or mild
soap
– Wash at 30°C in
the washing
machine (without
spinning)
Disinfection
Wipe disinfection
Not permitted
Wipe disinfection
Not permitted
Clean at 95°C, as
instructed in the
Immersion
disinfector
disinfection
instructions. Dry
thoroughly
Immerse in
Washing cycle up to
®
gigasept FF
(new)
95°C
(1)
solution
Immersion
Washing cycle up to
(1)
disinfection
95°C
Immersion
Washing cycle up to
(1)
disinfection
95°C
Immersion
Washing cycle up to
(1)
disinfection
95°C
Immersion
Washing cycle up to
(1)
disinfection
95°C
Thermo-
Sterili-
disinfector
zation
Not
permitted
Not
permitted
Steam ster-
ilization
Steam ster-
(2)
ilization
Steam ster-
(2)
ilization
Steam ster-
(2)
ilization
Steam ster-
(2)
ilization
Not
 (2)
permitted
(3)
(3)
(3)
(3)
(3)