Only operate the unit with a filter compartment cover or hygiene input
filter to prevent any liquids from entering the unit.
Always keep the air inlet openings on the filter compartment cover or the
suction inlets on the hygiene input filter clear.
Only have modifications to the unit carried out by the manufacturer,
WEINMANN Emergency, or by a technician expressly authorized by
WEINMANN Emergency.
Caution!
Do not place a switched-on cellular phone or radio closer than 1 m from
the MEDUMAT Transport, as this could cause malfunctions.
Remember that the respiratory resistance of the system as a whole may
increase beyond the level specified by the standard when an HME filter
(heat and moisture exchanger), a bacterial filter or a combined HME
bacterial filter is used. Please also follow the manufacturer's instructions
for use for the filter being used.
When operating the unit with the power supply unit, always connect the
unit to an easily accessible outlet so that it can be unplugged quickly in
the event of a malfunction.
When operating the unit with the power supply unit, make sure that the
power cord cannot cause anyone to trip or cause any obstruction. If
necessary, do not use an external power supply, but operate the unit
with the battery instead.
When operating the unit with the 12 V supply cord, always connect the
unit to an easily accessible vehicle electrical system receptacle so that it
can be unplugged quickly in the event of a malfunction.
When operating the unit with the 12 V supply cord, make sure that the
cord cannot cause anyone to trip or cause any obstruction. If necessary,
do not use the vehicle electrical system, but operate the unit with the
battery instead.
An alternative ventilation unit must be kept available in case a unit fails.
After using the unit in a dusty environment (e.g., a gravel plant), change
the suction filter (see "12.5 Changing the suction filter" on page 131) or
the hygiene input filter (see "12.6 Changing the hygiene input filter" on
page 132).
Only operate a unit with hygiene input filter with software version 6.1 or
higher.
Safety information