Chapter 9 Troubleshoot; Overview; Reading And Troubleshooting Fault And Warning Messages - Emerson Rosemount FCL Reference Manual

Free chlorine system with rosemount 56 transmitter
Hide thumbs Also See for Rosemount FCL:
Table of Contents

Advertisement

Reference Manual
00809-0200-3412
9
Troubleshoot
9.1

Overview

When the transmitter identifies a problem the word warning or fault appears
intermittently at the bottom of the display. To read a fault or warning message and
troubleshooting information, press INFO.
See
Use the diagnostic
Warning The instrument or sensor is usable, but you should take steps as soon as
Fault
9.2
Reading and troubleshooting Fault and
Warning messages
Complete the following steps to access Fault and Warning messages and determine what
they mean.
Procedure
1. With the main display showing, press INFO.
Rosemount FCL 56
feature.
possible to correct the condition causing the warning. Warning messages can
be turned off. To turn off warning messages, go to the main menu and choose
Display setup. Scroll to the Warning tab and turn off warning messages.
The measurement is seriously in error and is not to be trusted. A fault
condition might also mean that the transmitter has failed. Correct fault
conditions immediately. When a fault occurs, the analog output goes to 22.00
mA or the value programmed in Outputs. Fault messages cannot be turned
off.
The screen below appears.
Troubleshoot
May 2019
49

Advertisement

Table of Contents
loading

Table of Contents