Exchanging a Flow Cell

Exchanging a Flow Cell

For bio-inert modules use bio-inert parts only!
When
If an application needs a different type of flow cell or the flow cell needs repair.
Tools required
p/n
OR
5043-0915
Parts required
#
1
1
1
1
1
1
Preparations
Turn the lamp(s) off.
Remove the front cover.
Sample degradation and contamination of the instrument
C AU T I O N
Metal parts in the flow path can interact with the bio-molecules in the sample leading
to sample degradation and contamination.
172
Description
Wrench, 1/4 inch
for capillary connections
Fitting mounting tool
for bio-inert capillaries
p/n
G1315-60022
G1315-60025
G1315-60024
G1315-60015
G5615-60022
For bio-inert applications, always use dedicated bio-inert parts, which can be
identified by the bio-inert symbol or other markers described in this manual.
Do not mix bio-inert and non-inert modules or parts in a bio-inert system.
Description
Standard flow cell, 10 mm, 13 µL, 120 bar (12 MPa)
Semi-micro flow cell, 6 mm, 5 µL, 120 bar (12 MPa)
Micro flow cell, 3 mm, 2 µL, 120 bar (12 MPa)
High pressure flow cell, 6 mm, 1.7 µL, 400 bar (40 MPa)
Nano flow cell, refer to
"Nano Flow Cell - Replacing or Cleaning"
page 192
Standard flow cell bio-inert, 10 mm, 13 µL, 120 bar (12 MPa) for
MWD/DAD, includes Capillary Kit Flow Cells BIO (p/n G5615-68755)
Agilent 1260 Infinity DAD and MWD User Manual