The system conforms to the DICOM standard as specified in the SonoSite Edge II, SonoSite SII DICOM
Conformance Statement, available at www.sonosite.com. This statement provides information about the
purpose, characteristics, configuration, and specifications of the network connections supported by the
system.
Security
The port for DICOM communication (specified by the user in the system settings; typically port 104, 2762,
or 11112) is used for outgoing communication to the network.
Anti-virus software is not installed on this device.
This device has a single configurable listening port for the purposes of DICOM Echo and Storage
Commitment.
Data flow
DICOM
MWL Server------------------------> Edge II system-------------------------> PACS
Study order
(DICOM MWL)
Please refer to the SonoSite Edge II, SonoSite SII DICOM Conformance Statement (D18493) for details.
1 Connection of equipment to an IT network that includes other systems, could
Caution
2 When a setting of the IT network to which this device is connected has been
11-2
Study data
(DICOM Storage)
result in previously unidentified risks to patients, operators or third parties. Before
connecting the equipment to an uncontrolled IT Network, make sure that all
potential risks resulting from such connections, were identified and evaluated,
and suitable countermeasures were put in place. IEC 80001-1:2010 provides
guidance for addressing these risks.
changed, check that the change does not affect this device and take measures if
necessary. Changes to the IT network include:
Changes in network configuration (IP address, router etc.)
Connection of additional items
Disconnection of items
Update of equipment
Upgrade of equipment
Any changes to the IT network could introduce new risks requiring additional
evaluation to be performed as per item 1 above.