Print does not work
1 Select the printer on the Connectivity setup page. Refer to
page 3-9.
2 Check the printer connections.
3 Ensure that the printer is turned on and set up properly. Refer to the printer manufacturer's instructions,
if necessary.
System does not recognize the transducer
Disconnect and reconnect the transducer.
System prompts you to "ensure the USB device is valid."
Make sure that the USB storage device does not have password protection enabled and is not defective.
Use the USB storage device included with the system.
System prompts you to "ensure the USB device contains valid data."
1 Make sure that the data are present on the USB storage device.
2 Reexport the original data onto the USB storage device.
3 Contact your system administrator.
System displays the alert "Incompatible power supply . . . ."
Use the power supply that shipped with the system.
System displays the alert "The external video is not functional . . . ."
Make sure that the system is securely attached to the dock.
System displays the alert "Maximum number of procedure entries reached" when trying
to create a patient information form.
Free up internal storage space by archiving or exporting patient exams and then deleting them from the
system.
System displays the alert "Unable to save image or clip. You have reached the maximum
number of images/clips allowed for a single patient."
Delete any unwanted images or clips from the patient exam.
7-2
"To configure the system for a printer"
Troubleshooting and Maintenance