2 Do any of the following as needed:
Adjust controls as described in
Using the touchpad, position the D-line and gate where desired. Horizontal movements position the
D-line. Vertical movements position the gate.
PW Doppler
To correct the angle manually, do one of the following:
Press the SELECT key and then use the touchpad. The SELECT key toggles between the D-line
and angle correction.
Freeze the image, and then use the FAR gain knob to adjust the angle in 2° increments from -74°
to +74°.
To display the spectral trace
1 Display the D-line.
2 Press the D key. The time scale above the trace has small marks at 200 ms intervals and large marks at
one-second intervals.
3 Do any of the following as needed:
Adjust controls. as described in
Press the UPDATE key to toggle between the D-line and spectral trace.
If using a duplex layout, press the D key to toggle between the full-screen D-line and the duplex
layout.
To set a duplex layout, see
PW Doppler controls
Table 4-3: PW Doppler on-screen controls
Control
Description
PW, CW
CW applies to Cardiac exam only.
Toggle between PW Doppler and CW Doppler.
The current selection appears in the upper left-hand screen.
Corrects the angle to 0°, +60°, or -60°.
Angle
You may also adjust the angle using the touchpad.
Correction
Gate Size
Settings depend on transducer and exam type.
In TCD or Orb exams, use the touchpad to specify the Doppler gate depth (the depth of
the center of the gate in the Doppler image). The Doppler gate depth indicator is on the
lower right-hand screen.
Imaging
"PW Doppler controls"
"Spectral trace controls"
"Presets setup"
on page 3-14.
on page 4-7.
on page 4-7.