Medical Ultrasound Safety, American Institute of Ultrasound in Medicine (AIUM), 1994. (A copy is included
with each system.)
Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment, NEMA UD2-2004.
Acoustic Output Measurement and Labeling Standard for Diagnostic Ultrasound Equipment, American
Institute of Ultrasound in Medicine, 1993.
Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound
Equipment, NEMA UD3-2004.
Guidance on the interpretation of TI and MI to be used to inform the operator, Annex HH, BS EN 60601-2-37
reprinted at P05699.

Transducer surface temperature rise

Table 10-4
and
Table 10-5
transducers used on the ultrasound system. The temperatures were measured in accordance with EN
60601-2-37 section 42 with controls and settings positioned to give maximum temperatures.
Notes
Table 10-4: Transducer Surface Temperature Rise, External Use (°C)
Test
Still air
14.2
Simulated Use
7.3
Table 10-5: Transducer Surface Temperature Rise, Non-External Use (°C)
Test
ICTx
Still air
9.2
Simulated Use
5.2
10-10
list the measured surface temperature rise from ambient (23°C
For information about surface temperature rise for the P11x transducer, refer to
the P11x Transducer User Guide, included with the P11x transducer.
For information about surface temperature rise for the TEExi transducer, refer to
the TEExi Transducer User Guide, included with the TEExitransducer.
Transducer surface temperature rise is the same for the standard and armored
versions of a transducer.
The P11x transducer is not licensed for use in Canada.
15.3
15.0
12.4
8.5
8.9
7.7
C8x
11.3
5.5
10.7
17.5
16.1
7.7
9.1
8.5
±
3°C) of
12.5
16.0
14.9
8.8
9.1
7.6
Acoustic Output