Cleaning the stand
For information about cleaning the stand, see the Edge Stand and H-Universal Stand User Guide.
Cleaning accessories
Clean accessories prior to disinfecting. You can disinfect the exterior surface of accessories using a
recommended disinfectant. Refer to the cleaners and disinfection document available at
www.sonosite.com.
To clean and disinfect accessories
1 If necessary, unplug the power supply, and detach any cables.
2 Clean the exterior surfaces of the accessory using a soft cloth lightly dampened in a mild soap or detergent
cleaning solution to remove any particulate matter or body fluids.
Apply the solution to the cloth rather than the surface.
3 Mix the disinfectant solution compatible with the accessory, following disinfectant label instructions for
solution strengths and disinfectant contact duration. Refer to the cleaners and disinfection document
available at www.sonosite.com.
4 Wipe surfaces with the disinfectant solution. Follow the manufacturer's instructions for the disinfecting
solution, including exposure time and temperature for effective disinfection.
5 Air dry or towel dry with a clean cloth.
Cleaning and disinfecting the ECG cable and slave cable
Caution
To clean and disinfect the ECG cables (wipe method)
1 Remove the cable from the system.
2 Examine the ECG cable for damage such as cracks or splitting.
3 Clean the surface using a soft cloth lightly dampened in a mild soap, cleaning solution, or pre-moistened
wipe. Apply the solution to the cloth rather than the surface.
4 Wipe the surfaces with a FUJIFILM SonoSite Approved cleaner or disinfectant.
5 Air dry or dry with a clean cloth.
For more information on the ECG slave cable, refer to the ECG Slave Cable User Guide.
Cleaning and Disinfecting
To avoid damaging the ECG cable, do not sterilize.