Separation distance

The SonoSite Edge II ultrasound system is intended for use in an electromagnetic environment in which
radiated radio frequency (RF) disturbances are controlled. The customer or the user of the SonoSite Edge II
ultrasound system can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the SonoSite Edge II
ultrasound system as recommended below, according to the maximum output power of the
communications equipment.
Table 9-1: Recommended separation between portable and mobile RF communications
equipment and the SonoSite Edge II ultrasound system
Rated maximum
output power of
transmitter
Watts
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Notes

Compatible accessories and peripherals

FUJIFILM SonoSite has tested the SonoSite Edge II Ultrasound System with the following accessories and
peripherals and has demonstrated compliance to the requirements of IEC60601-1-2:2007.
Safety
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d=1.2 √P
0.12
0.38
1.2
3.8
12
At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects, and people.
80 MHz to 800 MHz
d=1.2 √P
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
d=2.3 √P
0.23
0.73
2.3
7.3
23
9-15