1 Quick Start
Function Inspection
1. Press Default to restore the oscilloscope to its default configuration.
2. Connect the ground alligator clip of the probe to the "Ground Terminal" under the
probe compensation signal output terminal.
3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the
"Compensation Signal Output Terminal" of the probe.
4. Press AUTO.
5. Observe the waveform on the display. In normal condition, the display should be
a square waveform as shown in the figure below:
6. Use the same method to test the other channels. If the square waveforms
actually shown do not match that in the figure above, please perform
Compensation" in the next section.
WARNING
To avoid electric shock, please make sure that the insulated wire of the
probe is in good condition and do not touch the metallic part of the
probe when the probe is connected to a high voltage source.
MSO4000/DS4000 User's Guide
Figure 1-7 To Use the Compensation Signal
Figure 1-8 Square Waveform
Compensation Signal Output Terminal
Ground Terminal
RIGOL
"Probe
1-9