5 To Trigger the Oscilloscope
the odd field follows behind.
PAL: the frame frequency is 25 frames per second. The TV sweep line is 625
with the odd field goes first and the even field follows behind.
SECAM: the frame frequency is 25 frames per second. The sweep line is 625
with interlacing sweep.
HDTV: HDTV consists of 480P, 576P, 720P, 1080P and 1080I display formats.
The specified video standards are as follows:
480P
576P
720P
1080P
1080I
Trigger Mode:
Press Sweep to select the
normal or single. The corresponding status light of the current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameters (trigger holdoff and noise rejection)
under this trigger type.
Trigger Level:
Use TRIGGER
description of
"Trigger
MSO4000D/S4000 User's Guide
the frame frequency is 60 frames per second. The TV sweep
line is 525. Line-by-line sweep. The line frequency is 31.5
KHz.
the frame frequency is 60 frames per second. The TV sweep
line is 625. Line-by-line sweep.
the frame frequency is 60 frames per second. The TV sweep
line is 750. Line-by-line sweep. The line frequency is 45 KHz.
the frame frequency is 60 frames per second. The TV sweep
line is 1125. Line-by-line sweep.
the frame frequency is 25 to 30 frames per second. The TV
sweep line is 1125. Interlacing sweep. The line frequency is
33.75 KHz.
Trigger Mode
LEVEL to modify the level. For details, please refer to the
Level" on page 5-9.
(page 5-3) under this trigger type as auto,