Rigol MSO4000 User Manual page 182

Digital oscilloscope
Hide thumbs Also See for MSO4000:
Table of Contents

Advertisement

RIGOL
3. MISO Setting
Press MISO to enter the MISO data line setting interface.
Press Channel to select any channel (CH1-CH4 or D0-D15) as the MISO
data channel. When "None" is selected, this data line is not set.
Press Polarity to set the polarity of the MISO data line to
level is 1) or
Press Threshold to set the threshold of the MISO data channel.
4. MOSI Setting
Press MOSI to enter the MOSI data line setting interface.
Press Channel to select any channel (CH1-CH4 or D0-D15) as the MOSI
data channel. When "OFF" is selected, this data line is not set.
Press Polarity to set the polarity of the MOSI data line
is 1) or
Press Threshold to set the threshold of the MOSI data channel.
5. Data Bits Setting
Press Data Bits to set the number of bits of each frame of data. The range
available is from 4 to 32.
6. Endian Setting
Press Endian to select "LSB" or "MSB" and the default is "MSB".
7. Display-related Setting
Press Format to set the display format of the bus to Hex, Decimal, Binary or
ASCII.
Press Offset and use
Press BusStatus to turn the bus display on or off.
8. Decoding Table
The decoding table displays the decoded data, the corresponding line number,
time and error information on the MOSI or MISO data line in table format. This
can be useful when observing longer data transimissions by presenting the data
in a tabular format.
Press Event Table  Event Table to select "ON" (note that this operation is
only available when BusStatus is set to "ON") to enter the decoding table
8-14
(the low level is 1).
(the low level is 1).
to adjust the vertical display position of the bus.
8 Protocol Decoding
(the high
(the high level
MSO4000/DS4000 User's Guide