P122: Return-to-Zero
Selects a weight threshold as a percentage of full scale below which the
gross weight must fall before another accumulation can be performed.
P124: Count Resolution
Selects the internal count resolution used in determining the quantity.
Normally this is set to zero (0) which uses the maximum internal resolution
when determining the quantity. However, when displaying a very large
count of very light pieces, the display may appear unstable. Decreasing
the count resolution will increase the count division size, thus making the
count appear more stable.
P125: Count Adjustment Factor
Assigns a conversion factor to the number of additional pieces that can be
added and still ensure the required accuracy is met. For example, if the
adjustment factor is 1.0 and the display reports you can add up to 200
additional pieces after performing a sample, changing the adjustment
factor to 0.1 would allow you to add up to 2000 pieces – changing the
adjustment factor to 10.0 would only allow you to add up to 20 pieces.
P126 – P130: Multi-Range
Configures the operation of the multi-range feature (see page 5-17).
P131 – P134: Units
Assigns the units selectable via the [UNITS] key from the weigh mode.
The units assigned at P131 will become the power-up units for the scale
presently specified at P108. Three additional units can be assigned at
P132 à P134 for access via the [UNITS] key.
P135: Rate Measurement Period
Assigns the rate measurement period (RMP) over which the rate of weight
change is averaged. For example, an entry of 2.5 will result in a 2.5
second rate averaging period. Thus, every 1/60
conversion rate) the rate reported at 23P will be updated to reflect the
average rate over the last 2.5 seconds (150 readings). A longer the RMP
will yield a more stable the rate display but will be slower to respond to
rapid changes in rate.
P136: Rate Time Unit
Sets the time measurement criteria for calculating the rate (i.e.
rate/second, rate/minute, rate/hour).
P142: Center-of-Zero Annunciator
Enables the center-of-zero annunciator. In multi-scale applications,
disabling the center-of-zero annunciator will allow you to view the scale
number when the weight is at center-of-zero.
S
TATUS
The status parameters allow renaming of the status word transmitted when
using operating parameter 97P.