Audit Trail Parameters; Calibration Parameters - GSE 60 SERIES Technical Reference Manual

Programmable process controllers
Table of Contents

Advertisement

14-8 Chapter 14
Table 14-4: Audit Trail Parameters
A
T
P
UDIT
RAIL
ARAMETER
P6020!
OIML
00001
P6020#
Cal.
00001
P6020$
Setup
00001
P6020%
MUST!
Check
Table 14-5: Calibration Parameters
C
P
ALIBRATION
ARAMETER

Calibration Parameters

P6109(
Scale
P6110)
Crrnt
mv/v
P6110!
CAL
Factr
P6110@
ReZro
Wght
P6110#
ZrTrk
Wght
P6110$
CZero
GSE Scale Systems
A
UDIT
Several audit trail parameters are used to satisfy the requirements of
various weights & measures agencies for sealing an instrument in legal-
for-trade applications. These are non-resettable parameters that
increment each time certain information is changed within the setup mode.
Table 14-4 shows the different audit trail parameters. Refer to chapter 5
for more information on legal-for-trade applications.
Displays the OIML audit trail number.
Displays the calibration audit trail number. This number increments each time a
new calibration is saved.
Displays the setup audit trail number. This number increments each time setup
changes are saved when exiting the setup mode.
Displays a list of parameters which, if configured improperly, could facilitate fraud in
a legal-for-trade installation. A weights & measures inspector may check this
parameter and inquire about the setup of any parameter that appears in this list.
C
ALIBRATION
The calibration parameters contain data internally calculated as the result
of the load cell, analog, and A/D converter calibration. These values are
unique to each instrument and should be recorded so they can be restored
in the event they are inadvertently deleted from the setup. With few
exceptions, calibration parameter values described in Table 14-5
should never be changed from the original values calculated during
initial calibration.
Used to enter the scale number for P61110 à P61121. For example, key in 2
[ENTER], then access P61110 à P61121 to view calibration data for scale #2.
1
Displays the current mV/V output of the load cell. The range is –5.0 mV/V to +5
mV/V with an accuracy of 0.02%. For example, a 100 lb capacity load cell rated at
3.00 mV/V at full scale should display 1.5 mV/V with 50 lb applied.
Displays the calibration factor calculated during the last load cell calibration. This
parameter does not appear with linearization enabled (P119). Refer instead to
P61130 à P61139 in this table for information on linearization values.
Displays the amount of weight (in default units per P150) that has been zeroed
through use of the [ZERO] key since the last calibration.
Displays the amount of weight (in default units per P150) automatically tracked off
by zero tracking since the [ZERO] key was last pressed.
Displays the coarse zero value calculated during the last load cell calibration.
0%
T
P
RAIL
ARAMETERS
D
ESCRIPTION
P
ARAMETERS
D
ESCRIPTION